Table 3 Accuracy of self-injectable epinephrine device demonstration (count and %)

From: Demonstration of epinephrine autoinjectors (EpiPen and Anapen) by pharmacists in a randomised, simulated patient assessment: acceptable, but room for improvement

Demonstration performed

Original EpiPen

New-look EpiPen

Anapen

Total

P value*

 

(n = 83)

(n = 87)

(n = 80)

(n = 250)

 

Step 1

69 (83.1)

76 (87.4)

75 (93.8)a

222 (86.8)

0.11

78 (97.5)b

Step 2

80 (96.4)

84 (96.6)

76 (95)

240 (96.0)

0.86

Step 3

62 (74.7)

59 (67.8)

61 (76.3)

182 (72.8)

0.42

Step 4

8 (9.6)

26 (29.9)

18 (22.5)

52 (20.8)

0.005

Steps 1-3

50 (60.2)

54 (62.1)

59 (73.8)

163 (65.2)

0.15

Steps 1-4

8 (9.6)

21 (24.1)

17 (21.3)

46 (18.4)

0.04

  1. *Pearson chi-squared p-value for comparison of demonstration accuracy across groups. Step 1: Remove safety caps; Step 2: Place against mid-anterolateral thigh; Step 3: Push down hard/press red button to inject and hold for 10 seconds; Step 4: Remove device, avoid needle, massage site. aRemove black needle shield. bRemove gray safety cap from red button (Anapen only).