Skip to main content

Table 6 Volume depletion per test for allergy skin test devices

From: Proceedings of the Canadian society of allergy and clinical immunology annual scientific meeting 2015

 

0.3 mL

1.0 mL

µL/Test

% CV

Tests/5 mL

µL/Test

% CV

Tests/5 mL

Duotip-test II®-gravimetric

2.89

15.6

1730

9.09

8.3

550

Duotip-test II®-volumetric

2.60

3.8

1923

9.15

0.9

546

Unitest PC®-gravimetric

10.22

8.4

489

20.10

4.9

249

Unitest PC®-volumetric

10.28

3.2

486

21.20

0.8

236

Multi-test II®-volumetric

7.15

3.3

699

9.92

6.9

504

Multi-test PC®-volumetric

8.02

1.5

623

13.82

3.2

362

Prick Lancetterâ„¢-drop

35

9.4

143

35

 

143