Table 4 Falls in percentage from 100% oxygen after five actuations of pMDIs into four ventilator delivery devices

From: Asthma pressurised metered dose inhaler performance: propellant effect studies in delivery systems

 

5× Duolin™

45 ml (O2 %)

5× Respigen™

60 ml (O2 %)

5× Vannair™

45 ml (O2 %)

5× Intal™

67.5 ml (O2 %)

L-Trace™

88

87

89

81

Spirale™

86

87

89

82

Intersurgical™

87

86

87

79

RT 200 (ventilator tubing)

90

91

93

91